← All technology

EASYRET

EASYRET

Abstract

The EasyRet is a 577nm yellow wavelength laser that uses proprietary ELBA fiber laser technology, delivering a uniform top-hat laser spot profile ideal for ophthalmic applications.

TECHNICAL SPECIFICATIONS

Laser Technology

  • Laser Source: Fiber laser technology (ELBA™)
  • Wavelength: Yellow 577 nm
  • Power at Tissue: 50-2000 mW maximum
  • Beam Profile: Top-hat (perfectly homogeneous laser spot profile)

Pulse Parameters

  • Pulse Duration: From 10 ms to continuous
  • MultiSpot Mode: Short pulse durations from 10 to 20 ms
  • SubLiminal® Mode: Microsecond pulses with customizable duty cycle adjustable from 5% to 100%

Spot Size Parameters

  • SingleSpot Mode: Continuously adjustable from 50 µm to 400 µm
  • Pattern Mode: Continuously adjustable from 100 µm to 400 µm

Aiming Beam

  • Wavelength: 635-650 nm (red)

Cooling System

  • Peltier effect cooling

Physical Specifications

  • Dimensions: 1742 × 970 × 720 mm
  • Weight: 60 kg
  • Power Requirements: 100 to 240 VAC, 350 VA, 50/60 Hz

Slit Lamp Integration

Available with two slit lamp types:

  • Haag Streit Type: Quantel Medical (CSO 990 5x)
  • Zeiss Type: Quantel Medical (CSO 980 5x)

Both feature:

  • Integrated laser adapter with continuously variable parfocal zoom
  • Large touch screen interface

TREATMENT MODES

1. SingleSpot Mode

  • Single point, repeat, pattern, and continuous modes
  • Traditional single laser spot delivery

2. MultiSpot Pattern Scanning Mode

  • Four customizable patterns: Single Spot, Squares, Triple Arcs
  • Delivers simultaneous multiple targets
  • Reduces treatment time (full PRP in 1 session)

3. SubLiminal® Subthreshold Mode

  • Converts each laser shot into a pulse envelope with customizable train of short pulses
  • Allows operator to fully adjust pulse duration (On Time) and interval (Off Time)
  • Non-visible laser impacts for tissue-sparing treatments

CLINICAL APPLICATIONS & MEDICAL INDICATIONS

Retinal Pathologies Treated

  • Diabetic Retinopathy (DR)
  • Diabetic Macular Edema (DME)
  • Proliferative Diabetic Retinopathy (PDR) - Pan-Retinal Photocoagulation (PRP)
  • Macular Edema
  • Central Serous Chorioretinopathy (CSC)
  • Peripheral retinal pathologies
  • Macular pathologies

Wavelength Advantages (577nm Yellow)

  • Excellent combined absorption by both melanin and oxyhemoglobin (peak absorption of oxyhemoglobin)
  • Very little absorption by macular xanthophyll pigments
  • Excellent penetration through cataracts and hazy media

Clinical Benefits

MultiSpot Mode:

  • Less heat diffusion to the retina and choroid
  • Less damage to the retinal nerve fiber layer
  • More comfortable treatment better tolerated by patients
  • Treatment time reduction

SubLiminal® Mode:

  • Tissue-sparing for macular treatments
  • Subthreshold treatment (non-visible laser impacts)
  • Precise thermal effect management
  • Reduces burden of monthly visits and costly injections for DME patients

ADDITIONAL FEATURES

Resume® Function

  • Exclusive function offering operators more flexibility in implementing treatments
  • Allows restarting or stopping a new pattern in case of interruption

Software Interface

  • Clinically oriented, intuitive user interface
  • Software-guided treatment procedures
  • Three dedicated aiming beams for each treatment mode
  • Customizable laser treatment reports

OPTIONAL MOSAR® IMAGING SYSTEM

  • HD imaging system
  • Co-observation teaching mode for live viewing
  • Library mode for managing images, videos, and treatment reports
  • USB drive and local network export capabilities

REGULATORY & WARRANTY

  • CE Marked (Europe)
  • FDA Approved (United States - July 25, 2017)
  • Warranty: 5-year guarantee on laser cavity

This represents one of the most advanced and versatile retinal photocoagulation platforms available, combining three distinct treatment modalities in a single integrated system.