Compact Touch


Abstract

The Quantel Medical Compact Touch® represents a significant advancement in portable ophthalmic diagnostics, combining A-scan, B-scan, and pachymetry capabilities in a single, compact device designed for modern eye care practices.

Advanced Imaging Technology

At the core of the Compact Touch is its next-generation 15MHz probe, delivering 30% improved resolution compared to previous models. This enhancement establishes new standards in B-mode imaging quality, enabling clinicians to visualize ocular structures with exceptional clarity for more accurate diagnoses.

Precision Pachymetry Specifications

The system's pachymetry function utilizes a 20MHz transducer with a 1.2mm tip diameter, measuring corneal thickness from 200 to 999 microns with ±5 micron precision. The contact method employs a 45-degree angle with convergence at 0.5mm from the tip, allowing 1-10 measurements per session. This precision is crucial for glaucoma monitoring and IOL power calculations.

Key Features and Capabilities

The Compact Touch offers comprehensive diagnostic tools including cartographic mapping and IOP correlation tables for advanced corneal analysis. The A-scan provides precise axial length measurements, while B-scan delivers real-time cross-sectional imaging for posterior segment evaluation.

The device features a modern touch screen interface and operates silently without fans, comparable to tablet operation. Its compact design fits any examination room while maintaining full diagnostic functionality.

Connectivity and Integration

Modern connectivity options include DICOM compatibility, WiFi and Bluetooth connectivity, and HDMI video output. This comprehensive suite enables seamless integration with PACS systems, EMR platforms, and existing network infrastructure, streamlining workflow and data management.

Clinical Applications

The Compact Touch addresses critical needs in contemporary practice through its portability and versatility. It's particularly valuable for multi-location practices, mobile clinics, and busy offices requiring flexible diagnostic capabilities. The precision pachymetry is essential for modern glaucoma management, where accurate corneal thickness measurements impact IOP interpretation.

Conclusion

The Quantel Medical Compact Touch successfully combines advanced ultrasound technology with practical clinical utility. By integrating three diagnostic modalities into a portable, high-performance platform, it offers unprecedented flexibility without compromising accuracy. For practices seeking comprehensive ophthalmic diagnostics in a versatile, space-efficient solution, the Compact Touch delivers cutting-edge technology designed for modern clinical environments.